Unique integrated electric pump for one handed pressure calibration
Test pressure switches easily with easy-to-use switch test function
Precision pressure adjust vernier for easy and accurate pressure calibration
Ideal accuracy and performance (0.025 %) for reference class transmitter calibration
30 PSI, 150 PSI and 300 PSI pressure ranges
Programmable pressure calibrator pump limit:
Eliminates over-pressurization
Allows the user to program target pressures for pumping
Extended pressure measurement range with any of 29 700Pxx Pressure Modules
Pump cleanout ports reduce pump failures from moisture and fluid damage
Fluid trap included with the unit to mitigate chance of fluid damage in at-risk applications
Precision temperature measurement with optional RTD probe
型號 描述 Fluke-719PRO-150G Electric Pressure Calibrator
Includes:
Fluke-719Pro-30G Electric Pressure Calibrator, 30 PSIG
Fluke-719Pro-300G Electric Pressure Calibrator, 300 PSIG
通用配件:
配件 描述 Fluke 720RTD RTD Probe for 721 and 719Pro
Fluke-71XTrap Liquid and Dirt Trap for 718 and 719Pro
Fluke-720URTDA Universal RTD Adapter for Fluke 721 and 719Pro
Specifications
Pressure Measurement
Range
-12 to 300 PSI / -0.8 to 20 bar Resolution
0.01 PSI, 0.001 bar Accuracy (1 year)
± 0.025 % of full scale, 6 months ± 0.035 % of full scale, 1 year Dry Air and non-corrosive gas only Temperature effect (all ranges)
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C No effect on accuracy on all functions from 15 °C to 35 °C Function
Range
Resolution
Accuracy (1 year)
Comment
mA dc (simulation and measurement) 0-24 mA dc 0.001 mA ± 0.015 % of reading ± 2 counts 1000 ohm maximum load in mA source, 26 V dc maximum voltage in mA simulate Volts dc (measurement only) 30 V dc 0.001 V ± 0.015 % of reading ± 2 counts No AC voltage measurement, do not exceed 30 V dc Temperature (measurement only, Pt100 RTD) -50 to 150 °C (-50 °F to 302 °F) 0.01 °C 0.01 °F Temperature ± 0.1 °C (0.2 °F) ± 0.25 °C (± 0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory) Loop power supply 24 V dc N/A 24 mA at 24 V Temperature effect (all functions) Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C No effect on accuracy on all functions from 15 °C to 35 °C Environmental
Operating temperature -10 °C to +50 °C Storage temperature (with batteries) Per battery manufacturer's specification within storage specification below Storage temperature (without batteries) -20 °C to +60 °C Power Requirements
Battery Eight (8) standard Alkaline AA batteries Battery life 300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 PSI; 100 pump cycles to 300 PSI minimum Physical
Dimensions (HxWxD) 24.13 cm x 11.18 cm x 7 cm (9.5 in x 4.4 in x 2.8 in) Weight With holster 1.2 kg (2.5 lb) Connections
Pressure 1/8 inch female NPT Electrical Standard banana jacks RTD 4 pin miniature circular External module 6 pin miniature circular Engineering Units (Configurable)
psi, bar, mbar, kPa, kgcm2, cmH2O@4 °C, cmH2O@20 °C, mH2O@4 °C, mH2O@20 °C, inH2O@4 °C, inH2O@20 °C,inH2O@60 °F, mmHg@0 °C, inHg@0 °C, ftH2O@4 °C, ftH2O@20 °C, ftH2O@60 °F Electric Pressure Calibrator, 150 PSIG
Test leads with alligator clips
Printed user's manual and user's manual on CD-ROM in 14 languages
Fluid trap
Carrying case
1/4 inch NPT to 1/4 inch ISO adapter
Accessory hose kit with (2) 1/8 inch quick fit adapters and 1 meter of hose
Traceable certificate of calibration